Sirona Dental Systems GmbH
5
Programming the Teleradiograph
1.
I6.0
80
I4
3.
41 63 817 D3297
D3297.201.04.07.02
30
40
CCD CEPH
CCD CEPH
18,40 x 0,59 cm
18,40 x 0,59 cm
CEPH-A
CEPH-S
SID = 166 cm
SID = 166 cm
2.
4.
5 Programming the Teleradiograph
The four patient symbol buttons are preprogrammed
with kV/mA combinations.
However, the buttons can be freely programmed with
other values.
"Program values - teleradiography" section,
see page 22.
Programming procedure
1.
Set diaphragm No. 30 or No. 40 on the diaphragm
wheel for the ceph function. On the Multitimer, dis-
play C above s illuminates.
2.
Press the - + buttons to set the desired kV and mA
values on the digital display.
3.
Press the memory button. The LED above the mem-
ory button flashes.
4.
Press the patient symbol button to be repro-
grammed. The LED above this button illuminates.
The LED above the memory programming button extin-
guishes. The new values are now stored.
Programming is complete.
Enter the newly-programmed values in the "Freely pro-
grammed values" table for teleradiographs (see page
22).
Adjusting the exposure
See the "Programming" chapter of the ORTHOPHOS
Plus DS Operating instructions.
21