Program Values - Tele-Radiography; Program Values; And P.a. Exposures; Radiography Of Carpus - Sirona ORTHOPHOS Plus DS Ceph Notice D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

6 Program Values – Tele-Radiography
6
Program Values – Tele-Radiography
6.1

Program Values

Factory-programmed values
73
73
kV
mA
15
15
Possible kV/mA combinations
60
60
60
60
9
10
12
14
6.2
a.p. and p.a. Exposures
Encrease kV/mA-values for a.p. and p.a. exposures by 2 steps
(at 84 kV/13 mA to 90 kV/12 mA).
6.3

Radiography of Carpus

Recommended value for radiography of carpus: 64 kV/16 mA
22
77
84
14
13
60
62
64
66
16
16
16
16
The beam time is constantly max. 16s and cannot be
changed.
Comparable to a conventional X-ray film exposure with
a sensitivity class of 200, this means an exposure time
of approx. 0.4s.
Freely programmed values
kV
mA
69
71
73
77
15
15
15
14
Sirona Dental Systems GmbH
– please enter here –
80
84
90
14
13
12
mA
41 63 817 D3297
D3297.201.04.07.02
kV