Fh Adjustment Help Message - Sirona ORTHOPHOS Plus DS Ceph Notice D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

4 Operating
12
4.2.I
FH
19
42I
FH
19
CCC
FH

FH adjustment help message

If the Ready LED keeps flashing and Help LED (19) for
height adjustment on Control Panel A is still illumi-
nated, this can mean that the most recently taken telera-
diograph was terminated prematurely.
The secondary diaphragm on the head support is then
not in the FH position.
This is signalled by the dots under the three digits on the
height adjustment display.
To move the head support to the FH position, press and
do not immediately release one of the buttons for height
adjustment ↑ or ↓.The dots then disappear. Help LED
(19) extinguishes.
In this case, the system performs a movement in oppo-
site directions so that the fixing devices at the head sup-
port remain in the previously set position. Consequently,
the patient may be in the head support.
i
NOTE
Please ensure that, with travel into the FH position, the
movement in opposite directions is executed in any
event. Otherwise, let the patient step out of the head sup-
port.
FH
ATTENTION
If is shown in the height adjustment displayed
the head support runs into the FH position without per-
forming a compensating movement, when keys ↑ or ↓ on
Control Panel A are actuated.
ATTENTION
Under no circumstances must the patient then be in the
head support!
Sirona Dental Systems GmbH
41 63 817 D3297
D3297.201.04.07.02

Publicité

Table des Matières
loading

Table des Matières