Chapter 5: Error Messages From Operating Light Indicators - NSK iCare+ Mode D'emploi

Masquer les pouces Voir aussi pour iCare+:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

CHAPTER 5: ERROR MESSAGES FROM OPERATING LIGHT INDICATORS

Description of operating light indicators:
The different indicators and the Control panel are useful to inform the operator in real time about the operating
conditions of the iCare+ device.
A
D
A. "Cle" indicator informs the operator that n.Clean product level is too low.
iCare+ device stops. It is necessary to replace the bottle by a new one in order to restart the device. Remove
the used bottle with a firm upward movement. To install a new bottle refer to paragraph 3.2.1.1.
B. "Cid" indicator informs the operator that n.Cid product level is too low.
iCare+ device stops. It is necessary to replace the bottle by a new one in order to restart the device. Remove
the used bottle with a firm upward movement. To install a new bottle, refer to paragraph 3.2.1.1
C. "Oil" indicator informs the operator that the lubrication Oil product level is too low.
iCare+ device stops. It is necessary to fill the oil tank. To fill the oil tank, refer to paragraph 3.2.1.2.
After changing a bottle (n.Clean or n.Cid) or filling the Oil tank, press "Start" to prime the circuit.
D. "Drawer" indicator informs the operator about 2 things:
Either the drawer is not properly positioned (or even absent). In this case, check that the drawer is
present and correctly inserted.
Or the drawer is full and might overflow. In this case, remove the drawer and empty it.
In both cases, the iCare+ is unable to start the cycle.
E. This light indicator informs the operator that there is a problem with the air pressure (too low air pressure
value, no air and too high air pressure value). Air pressure plays an important role in the cycle efficacy;
iCare+ device cannot start the cycle until the problem is not solved.
F. This light indicator informs the operator that the door is open or not properly closed. Close the door to start
any cycle.
iCare+ User Manual V11.4 • 09/2017
B
C
E
F
p. 36

Publicité

Table des Matières
loading

Table des Matières