CHAPTER 6: TRACEABILITY
The cleaning and disinfection phases have to be performed in accordance with the different current
recommendations. It is therefore necessary to enable the validation of the handpieces treatment done by the iCare+
device. That's why traceability software must be used. The software "SoftCare+" was created to ensure the
traceability monitoring of the cycles realized by the device.
Please refer to specific user manual of annex software destined to visualize cycles information.
CHAPTER 7: TROUBLE SHOOTING
PROBLEM(S)
The device does not switch on
The device does not clean and/or
lubricate
The device does not clean and/or
lubricate correctly
iCare+ User Manual V11.4 • 09/2017
iCare+ is equipped with a microprocessor that checks in real time every internal
component. If any problem is detected on one of the component, iCare+ device stops
immediately
POSSIBLE CAUSE(S)
Is the supply cord properly
connected to mains?
Is the supply cord properly
connected to the iCare+ device?
Is the switch positioned on the
"ON" position?
Are the fuses working?
Are the mode light indicators lit?
Is the door opened? (door LED lit)
Are the n.Clean or n.Cid level LEDs
lit?
Is the Oil level LED lit?
Is the air supply tube properly
connected to the device?
Is the air supply pressure set
between 5 and 6 bar?
Is Short mode selected?
Is the instrument properly
connected to the connector?
SOLUTION(S)
Connect the supply cord to
mains
Insert properly the cord plug in
the device's main connector
Turn the main switch to "ON"
position
Change the fuses with the
same model
Push on one of the selection
cycles buttons
Close the door properly
Change the corresponding
bottle and prime
Fill the Oil tank
Connect properly the air supply
tube to the device's connector
Set the air supply pressure
between 5 and 6 bar
Select Long mode
Plug the instrument properly
p. 37