Error Codes
If an error code appears on the LCD panel, see the following table.
Error code
Alarm audio volume
setting at the last
operation was not
stored.
An incompatible probe
is connected to the
control unit.
Power source failure
at the measuring part.
Communication failure
with the measuring
part.
*Error indication is reset by turning off the power.
13
Troubleshooting
When a problem is detected, check the following again before requesting a repair.
If none of these is applicable or if the trouble is not remedied even after an action
has been taken, a failure of this product is suspected. Contact your Authorized NSK
Dealer.
Trouble
The power does
not turn on.
16
Error
Cause
Dry cell batteries are not
inserted.
Dry cell batteries are not
inserted correctly.
The remaining battery level
is low.
Cause
The battery voltage
dropped when the
alarm audio volume
was set.
The control unit
cannot recognize the
probe.
The voltage exceeded
the voltage range at
the measuring part.
The measuring part is
not operating.
Insert dry cell batteries.
Correctly insert the batteries.
Replace with new dry cell batteries.
Remedy
If the battery remaining
level is low, replace the
batteries.
Connect a correct
probe.
Check if the probe is
securely connected.
If the battery remaining
level is low, replace the
batteries.
If the battery remaining
level is low, replace the
batteries.
Remedy