Troubleshooting

Error code

Type of error
Calibration error�
CE0
Beyond the upper
limit�
Calibration error�
CE1
Below the lower
limit�
Abnormal heat
E-F
generation from
the battery
8 – 2

Problems and actions

When trouble is found, check the following again before contacting your dealer� If none of these is
applicable or the trouble is not remedied even after action has been taken, a failure of this product is
suspected� Contact your dealer�
[Control unit, AC adaptor]
Problem
The power is not
turned on�
AC IN lamp does
not light�
The operating life of the motor
handpiece or contra angle has
expired�
The battery generates
abnormal heat�
Cause
The AC adaptor is not connected�
The plug of the AC adaptor is not
inserted into the outlet, or there is
no electricity in the outlet�
The battery is empty�
No battery is inserted�
The fuse has burnt�
The AC adaptor is not connected�
The plug of the AC adaptor is not
inserted into the outlet, or there is
no electricity in the outlet�
The fuse has burnt�
The AC adaptor's fuse has burnt�
Cause
Replace the motor handpiece or
contra angle head�
Replace the battery� If the heat
generates from the new battery,
malfunction of the circuit may be
suspected� Contact your dealer�
Check the connection�
Check the connection�
Charge the battery or use the AC adaptor�
Insert the battery, or use the AC adaptor�
Contact your dealer�
Check the connection�
Check the connection�
Contact your dealer�
Contact your dealer�