EXAMPLE OF CONNECTION
Cut off the air arrival. Connect
the tube coming from the unit
and from the iCare+ device on
the "T" connector. Connect
the compressor's arrival on
the "T" connector
iCare+ User Manual V11.4 • 09/2017
Following the norm EN ISO 7494-2, the air quality provided to iCare+ has to
be dry, clean and free from bacteria and contamination.
The pressure of the air supply should be between 5 and 6 bar
The minimum air flow level provided to the unit is 50 L./mn.
It is necessary to follow the above mentioned parameters in order to avoid
some dysfunctions or damage of the unit.
DENTAL UNIT
AIR COMPRESSOR
Figure 7: Connection of the iCare+ device to the air system
The pressure of the air supply must not be under 5 bar or above 6 bar. A
pressure sensor detects if the air pressure is too low or too high and stops
the machine. The operator is informed by an operating light indicator
positioned on the Control Panel.
Do not fold or deform the air supply tube
If you experience some difficulties during the connection step, please
contact your retailer
The air filter has an effective filter size of 5µm
AIR CONNECTION
iCare+ DEVICE
"T" CONNECTOR
p. 13