Figure 10: Implementation of
n.Clean bottle (500 mL)
Push the bottle of n.Clean solution into the correct housing (see Figure 10) and after that do the same thing with the
n.Cid solution bottle (see Figure 11). To insert the bottles, a short pressure on them is enough. It is not necessary to
push them strongly because they can be damaged.
iCare+ User Manual V11.4 • 09/2017
For each bottle of solution or for the Oil, a coloured sticker is used to avoid
confusions between the products during the implementation of them:
For the n.Clean solution, a yellow sticker on its bottle and a yellow safety
adhesive label on its housing (of the Top rear panel) are glued
For the n.Cid solution, a grey sticker on its bottle and a grey safety adhesive
label on its housing (of the Top rear panel) are glued
The oil is directly poured into the tank (on the Top rear panel) on which a
blue safety adhesive label is glued
Always use n.Clean and n.Cid bottles supplied by NSK Europe GmbH.
The efficiency of the iCare+ device is optimized and certified for n.Clean and n.Cid
products.
Cleaning and disinfecting protocols have been calculated using these products.
The use of any other product may lead to a failure or damage of the iCare+ device or
of the instruments connected on it, and the efficiency of the device will not be
ensured.
Therefore, in case of non respect, the user will be considered as responsible.
CAUTION: When the bottles are not inserted, the needles, if reachable, can cause a
risk of injury
Figure 11: Implementation of
n.Cid bottle (500 mL)
p. 16