Periodical Maintenance Checks; Error Code - NSK Varios Combi Pro Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

12

Periodical Maintenance Checks

Refer to the check sheet below and perform periodical maintenance checks every three months. Contact your Authorized NSK Dealer if any
abnormalities are found.
Points to Check
Ultrasonic Handpiece
Chamber Cap
Powder Spray
Water Spray
Chamber
Attaching/removing the Handpiece, Handpiece
Hose, Irrigation Bottle and Chamber

Error Code

13
"Error" and an Error Code are displayed on the LCD Panel of the Control Unit if this product stops operating due to malfunction, excessive
pressure, disconnection, or incorrect operation. Depress the Foot Control button D to perform the error check again. If the problem is not
detected, the error status is released and this product become operable. If the Error Code is still displayed, refer to the table below and take the
necessary steps to resolve the problem.
Error Code
Description of Error
0
System Error
1
24V voltage abnormality
2
12V voltage abnormality
3
Pump current abnormality
4
Air pressure abnormality
5
Heater voltage abnormality
6
Heater current abnormality
7
Heater temperature abnormality
8
Foot Control communication abnormality
9
Ultrasonic Handpiece abnormality
12
Touch sensor communication abnormality Turn the POWER switch to OFF and then to ON. If the problem still persists, contact
13
Error in self-check of the Ultrasonic Unit
14
Ultrasonic Unit system error
Operate the Ultrasonic Handpiece to confirm there is no abnormal vibration, noise, heat, etc.
Also check that the LED lights and water flows.
Check that there are no large/deep scratches.
After filling the powder chamber, spray powder into a spittoon, etc., and confirm that powder is
not leaking from connections other than the Nozzle Tip connection.
After filling the powder chamber, spray powder into a spittoon, etc., and confirm that water is
not leaking from connections other than the Nozzle Tip connection.
Check that there are no large/deep scratches.
Check that they can be attached and removed smoothly. If they are too loose or too tight, the
O-rings may be damaged or deteriorated. Replace O-rings. (Refer to "10 Care and
Maintenance.")
Turn on the Control Unit power again. Contact your Authorized NSK Dealer if the
error does not disappear.
Check the connection of the Irrigation Pump. If the problem still persists, contact
your Authorized NSK Dealer.
Check the incoming air pressure. If the problem still persists, contact your
Authorized NSK Dealer.
Turn on the Control Unit power again. Contact your Authorized NSK Dealer if the
error does not disappear.
Check the water circuit and supply water to the Control Unit.
Leave the Control Unit until it cools down before using it.
Contact dealer if the error does not disappear.
Check the connection of the Foot Control. If the problem still persists, contact your
Authorized NSK Dealer.
Check the connection of the Handpiece. If the problem still persists, contact your
Authorized NSK Dealer.
your Authorized NSK Dealer.
31
Details
Check/Remedy

Publicité

Table des Matières
loading