4.2.6 Hygienic cleaning of the holder for external canister "Bag"
Please note the respectively applicable hygiene instructions. Unless otherwise direc-
ted, please follow these steps:
1. Rinse the holder for external canister "Bag" under running water.
2. Place the holder for external canister "Bag" in a water bath at 65 °C
temperature for 15 minutes once a day.
3. Subsequently rinse the holder for external canister "Bag" thoroughly and allow
it to dry.
Löwenstein Medical GmbH & Co. KG recommends to replace the holder for external
canister "Bag" every 12 month, if it is used and cleaned frequently and for each
new patient.
4.2.7 Hygienic cleaning and disinfection of the suction tube
1. Dispose of the suction tube properly before the device is used by a new patient!
2. If the device is used for one patient only, rinse the suction tube with clear water
(max. 0.5 l) using the rinsing bottle after each aspiration and place it in a water
bath at 65°C temperature for 15 minutes once a day.
3. Afterwards, rinse the suction tube thoroughly with clear water and let it dry.
Even when the suction tube is used for one patient only, Löwenstein Medical GmbH
& Co. KG recommends to replace the suction tube at least every 4 weeks and
for each new patient since the material may become discolored and brittle due to
frequent applications!
4.2.8 Hygienic cleaning and disinfection of the rinsing bottle (250 ml)
1. Remove the lid of the rinsing bottle and rinse both parts under running water.
2. Place the rinsing bottle and the lid in a water bath at 65°C temperature
for 15 minutes once a day.
3. Afterwards, rinse the components thoroughly and let them dry.
4. Place the lid back on the rinsing bottle.
Löwenstein Medical GmbH & Co. KG recommends to replace the rinsing bottle
every 4 weeks, if it is used and cleaned hygienically frequently and for each new
patient!
4.3 Further use of the device
The secretion aspirator Allegra M30 is suitable for further use. However,
prior to passing the device on to other patients resp. persons, it has to be processed
professionally. For this purpose, please hand the secretion aspirator Allegra
M30 over to Löwenstein Medical GmbH & Co. KG or to qualified personnel au-
thorized by Löwenstein Medical GmbH & Co. KG. In this regard, please observe the
indications in section 6.1!
Maintenance
EN
36