Hygienic preparation

5.1 General information

5.2 Intervals

5.3 Hygienic preparation of the modules

Requirement
Part
Housing
Connection line
Wear suitable protective equipment for disinfection work.
Please refer to the instructions for use supplied with the
disinfectant used.
Interval
Weekly
As necessary, e.g., following infectious
diseases or severe soiling
Damage to the device caused by ingress of liquids!
Ingress of liquids may damage the therapy device, components,
and accessories.
 Do not immerse the therapy device, components, or
accessories in liquids.
The therapy device is switched off (see instructions for use for the
therapy device).
1. Remove the prismaPSG module from the therapy device
"4.2 Removing the prismaPSG module", page
2. Carry out hygienic preparation of the module as specified in
the following table:
Cleaning
Wipe with a damp cloth
using water or mild soap
Action
Clean the modules
Disinfect the modules
Disinfection
Wipe-down disinfection
(Recommendation: terralin
protect or perform advanced
Alcohol EP)
prismaPSG
5 Hygienic preparation
(see
32).
Sterilization
®
Not permitted