Use the socket for the 12 V power supply unit (fig. 1(D)) to connect the device to
the mains power supply with the aid of the supplied power supply unit to charge or
operate it as needed. Make sure that the device is positioned in such a way that it is
possible to easily disconnect it later.
Information on the permissible environmental conditions during operation can be
found in section 7 "Technical data".
suction tube
device with
secretion canister
power supply unit
(type: GTM91099-
6015-3.0-T2)
power outlet
Fig. 2
Connection of the secretion aspirator Allegra M30 to patient and accessories
aspiration catheter
fingertip
patient
mains cable
3.2 Startup
•
Take the device and the accessories out of the packaging.
•
During the aspiration procedure, ensure that the device is in an upright position
to prevent the secretion canister overflowing.
•
Prior to the initial startup it is essential to follow the safety instructions in section
1.6.
•
Only operate the device with inserted bacterial filter and connected secretion
canister resp. with the disposable secretion canister and the associated adapter.
Depending on the variants supplied, individual parts will already be inserted/
connected on delivery; note the following instructions for all further applications.
•
Always have one backup bacterial filter resp. additional disposable liner "One-
Way" at hand, since it is absolutely essential for safe operation!
•
Prior to each application, check that the reusable secretion canister is completely
clean to avoid frothing.