5
Hygienic preparation
5.1 General information
5.2 Intervals
5.3 Hygienic preparation of the module
Requirement
Part
Housing
Connection line
•
Wear suitable protective equipment for disinfection work.
•
Please refer to the instructions supplied with the disinfectant
used.
Interval
Weekly
As necessary, e.g., following infectious
diseases or heavy soiling
Damage to the device caused by ingress of liquids!
Ingress of liquids may damage the therapy device, components,
and accessories.
Do not immerse the therapy device, components, or
accessories in liquids.
The therapy device is switched off (see instructions for use of the
therapy device).
1. Remove the module from the therapy device
Removing the prismaCONNECT module", page
2. Carry out hygienic preparation of the module as specified in
the following table:
Cleaning
Wipe with a damp cloth
using water or mild soap
Action
Clean modules
Disinfect modules
Disinfection
Wipe disinfection
(Recommendation: terralin
protect or perform advanced
Alcohol EP)
prismaCONNECT
5 Hygienic preparation
(see "4.2
36).
Sterilization
®
Not permitted
EN
39