2.5 Information on the battery
Inspection of the internal power supply
•
Since the internal battery is not maintained automatically in a fully
operational state, it is necessary to check the charging status regularly
and, if necessary, the battery must be replaced by service staff.
•
The battery may only be replaced by trained service staff, since the
replacement by inadequately trained personnel could result in a
hazard (such as excessive temperatures, fire or explosion)!
The charge state of the battery is indicated by the corresponding LED status indica-
tor (for explanation see section 3.3.1).
Prior to the initial startup of the secretion aspirator Allegra M30 it is urgently
recommended to charge the battery completely and to repeat this after the first
uses.
The secretion aspirator Allegra M30 is equipped with a lithium ion battery
which only has a limited self-discharge on its disposal in contrast to conventional
battery types.
Lithium ion batteries do not have a memory effect. Therefore they can and should be
recharged at any time after initial charging. Only frequent short charging should be
avoided. The typical service life of the battery amounts to approximately 500 load
cycles, subsequently approximately 70% of the primary capacity are still available.
The battery of the secretion aspirator Allegra M30 is protected against
depth discharge by means of protective measures; however the instructions on
charging mentioned above should be observed. The battery is also protected against
overheating during charging. If the battery temperature is exceeded during charging
due to ambient conditions not intended, charging will be automatically interrupt-
ed temporary to allow cooling. This feature ensures the safety of the battery and
protects it from wear.
2.6 Warranty
WARNING:
Modifications of the ME device are not permitted.
The secretion aspirator Allegra M30 is covered by warranty for
2 years. No extension or renewal of the warranty is granted in the event that war-
ranty work is performed.
The battery is covered by warranty for 6 months. Wearing parts are excluded from
the warranty.
Löwenstein Medical GmbH & Co. KG is responsible for impacts on safety, reliability
and specific performance only if:
•
original accessories and replacement parts are used,
•
maintenance and repair are performed by professionals authorized by
Löwenstein Medical GmbH & Co. KG or by
Löwenstein Medical GmbH & Co. KG itself,
the secretion aspirator Allegra M30 is used and operated in
•
accordance with the instructions for use and within its intended use.
All warranty claims shall be void if:
•
the device is opened by unauthorized personnel,
•
the safety seal is removed/broken,
•
repairs are performed by unauthorized personnel,
•
modifications are made to the device,
since the basic safety of the device can no longer be guaranteed in these
cases and functional limitations may occur.