3.4.2 Secretion aspirator Allegra M30 with disposable secretion canister
Health risks due to the handling of infectious or pathogenic
germs
Infectious and pathogenic germs in the aspirate cause health risks.
•
Wear suitable disposable gloves by the decommissioning.
•
The reuse of bacterial filters for multiple patients is prohibited for
hygiene and safety reasons!
•
For each new patient, processing by Löwenstein Medical GmbH & Co.
KG or by an authorized service partner of Löwenstein Medical GmbH
& Co. KG is strictly required!
•
After each aspiration procedure, all components that have come into
contact with the aspirate must be cleaned, disinfected or disposed of.
•
The aspirate and the parts contaminated with aspirate must be
disposed of properly.
1. Switch off the device after the aspiration by holding the on / off button
(fig. 21(F)) pressed for 1-2 seconds.
2. Disconnect the power supply unit from the mains supply (100 V to 240 V AC),
if attached, and remove the device plug from the aspiration device afterwards.
3. Remove the suction tube (fig. 1.2(G)) from the patient connection (fig. 11(B)).
4. Clean or dispose of the suction tube properly (see section 4.1.7 resp. 4.2.7).
5. Remove the disposable liner "OneWay" of the secretion aspirator
Allegra M30 as described in chapter 3.3.4 up to and including point 5.
subsequently follow the hints regarding the cleaning / disinfection of the external
canister "Bag" in chapter 4.1.5 resp. 4.2.5 and the holder for the external
canister in chapter 4.1.6 resp. 4.2.6.
6. Clean the surface of the device according to section 4.1.2 resp. 4.2.2.
Recommendation in the event of an extended storage period: Keep the
device in suitable packaging, e.g. the shipping carton, until it is used the
next time.