7 Technical data
Model name
Allegra M30
Flow rate*
30 l/min ± 2 l/min (high flow)
(measuring point at
suction tube nozzle)
Vacuum
max. -90 kPa (~ -900 mbar; -675 mmHg)
(high vacuum)
Conversion factor: 1 kPa ~ 7.5 mmHg;
1 kPa ~ 10 mbar
Canister
Reusable secretion canister (1000 ml)
Disposable liner "OneWay" (1000 ml)
Suction tube
PVC, length 150 cm, vacuum stable up to
-0.9 bar
item number:
100713-2
inner diameter 6 mm, wall thickness 2 mm
100765
inner diameter 7 mm, wall thickness 1.5 mm
Nominal voltage of the
In:
power supply unit
Out: DC 12 V/5.0 A
Maximum load current
5.0 A
Permissible input voltage
12 V
Power consumption at 12 V
60 W
Degree of protection
Type BF
pursuant to IEC 60601-1
Risk classification pursuant to
IIa
93/42/EEC, IX
Protection class pursuant to
II
IEC 60601-1
AC 100 – 240 V ~ 50-60 Hz/1.5 A
IP degree of protection
IP22
CE marking
CE0197
Sound emission
without vacuum: approximately 60 dB(A)
Transport/Storage
Ambient conditions
ambient temperature: -25 °C to +70 °C
relative humidity:
up to 90 %, non-condensing
air pressure: 700 hPa to 1060 hPa
Operation
ambient temperature: +5 °C to +40 °C
relative humidity:
15 % to 90 %, non-condensing
air pressure: 700 hPa to 1060 hPa
Battery
10.8 V; 5800 mAh (lithium ion battery)
Charging time if battery is
~ 3 h
empty
Charging time if battery is
~ 2 h
approximately 50 % full
Power supply unit
GTM91099-6015-3.0-T2, GlobTek Inc.
Dimensions basic device
213 mm x 260 mm x 124 mm
(H x W x D)
Weight (basic device)
2.8 kg
Operating time
5 minutes ON; 5 minutes OFF
Runtime in battery
approximately 45 minutes,
operation
depending on the runtime of the motor
Lifetime
5 years
Item number (REF)
1510215hl
* The information provided may differ depending on the altitude above sea level, the
prevailing air pressure and the air temperature.