3 Operation
5. Connect the suction tube to the port in the middle of the secretion canister lid.
In doing so, ensure that the tube is fixed sufficiently tight.
Fig. 10
6. Connect the other ending of the suction tube to the aspiration catheter with the
aid of a fingertip.
EN
21
3.2.2 Connecting the diposable secretion canister
Malfunction due to collapsing disposable liner " OneWay"
A leak in the external canister "Bag" or at the lid of the disposable liner
"OneWay" may cause air to flow into the external canister "Bag". This
may lead to the collapse of the disposable liner "OneWay".
•
Inspect the disposable exudate canister system (1,000 ml) to ensure
that the lid of the disposable liner "OneWay" is firmly connected to the
external canister "Bag".
•
Ensure that all connections are firmly attached and properly connected.
•
Ensure that the external canister "Bag" is undamaged and the T-piece
is firmly attached.
•
Follow the instruction for use supplied by the manufacturer!
The original disposable secretion canister system consists of the external canister
"Bag", the holder for the external canister "Bag", the disposable liner "OneWay", the
connecting tube for the disposable liner "OneWay" and the sterile disposable suction
tube with step connector.
Connection designation
A
Vacuum connection
B
Patient connection
Fig. 11
Please also follow the instruction for use supplied with the disposable
secretion canister system (1,000 ml)!