Operation
3.3.4 Replacement of the disposable secretion canister
Health risks due to the handling of infectious or pathogenic
germs
Infectious and pathogenic germs in the aspirate cause health risks.
•
Wear disposable gloves in doing so when replace the disposable liner
"OneWay".
•
Use the the disposable liner "OneWay" for one patient only.
•
Replace the disposable liner "OneWay" incl. suction tube in accordan-
ce with the respectively applicable hygiene instructions, if it is full, prior
to each new patient or weekly at the latest.
•
For each new patient, professional hygienic processing by Löwenstein
Medical GmbH & Co. KG or an authorized service partner of Löwen-
stein Medical GmbH & Co. KG is strictly required!
•
Components that have come into contact with the aspirate must be
cleaned, disinfected or disposed of after each aspiration.
•
The disposal of aspirate and contaminated components must be
performed in a proper manner.
1. Switch off the secretion aspirator Allegra M30.
EN
2. Separate the suction tube from the patient connection.
27
3. Separate the pre-assembled connecting tube of the disposable liner " OneWay"
at the upper end of the T-piece of the external canister "Bag".
4. Remove the disposable liner "OneWay" from the reusable external canister
"Bag".
5. Dispose of the disposable liner "OneWay" incl. the suction tube in a proper
manner (Please refer to chapter 6.3 "Disposal").
6. Place a new disposable liner "OneWay" in the reusable external canister "Bag"
as specified in 3.2.2. Ensure that the connecting tube as well as the lid of the
disposable liner are properly seated on the external canister.
7. Attach a new suction tube to the patient connection of the disposable liner
"OneWay".
8. Switch on the secretion aspirator Allegra M30.