5 Troubleshooting

Malfunction
Device does not
start
Device does not
charge
Insufficient flow
rate
No flow rate
Device
switches off
Display
Probable cause
do not illuminate
Battery is discharged.
blinks red
Battery is discharged and impermissible input
voltage.
blinks red
Internal error
blinks red
Battery temperature too high for charging
and
illuminates green
Battery is defective.
blinks red
Impermissible input voltage
illuminates green/orange
Bacterial filter is blocked.
Leak in the suction line or in the canister
and/or
illuminates green
Service life of vacuum pump expired
continuous red light
Float sphere of the overflow protection blocks
the aspiration lid inlet.
Disposable liner "OneWay" is full
Bacterial filter is blocked.
Suction tube is blocked.
Bacterial filter is not inserted.
blinks red
Internal error
Remedy
Connect the power supply unit for charging or
operation.
Connect the correct power supply unit, check the
power supply.
Please contact service!
No measure required, automatic resumption
of charging when battery has cooled down.
Please contact service!
Connect the correct power supply unit, check the
power supply.
Replace the bacterial filter.
Check that secretion canister lid and tube
connections are tight and no leakages are
apparent.
Please contact service!
Check fluid level in the secretion canister,
empty the canister.
Replace the disposable liner "OneWay"
Replace the bacterial filter.
Replace the suction tube.
Insert the bacterial filter.
Please contact service!