3.3.2 Setting the vacuum
1. Switch on the device by pressing the on/off button (fig. 21(F)).
2. Select the desired vacuum range by pressing the corresponding button ("Low",
"Medium" or "High") on the control panel (fig. 21(A, B, C)).
As described in section 3.3.1, the vacuum setting buttons correspond to
certain vacuum ranges.
3. You can perform fine adjustments of the set vacuum via the additional air
opening on the fingertip (fig. 22).
Fig. 22
4. Switch off the device by holding the on/off button (fig. 21(F)) pressed for 1-2
seconds.
3.3.3 Aspiration
1. Perform the aspiration procedure as you were instructed by specialized person-
nel.
2. Control the aspiration procedure and the desired vacuum via the additional air
opening on the fingertip (fig. 22).
3. Decommission the device after the aspiration procedure according to section
3.4.
In any case, please take note of section 2.4 and the important information
on the handling of the bacterial filter! Ensure that the battery has a suffi-
cient capacity during the aspiration procedure. The battery status display
must signalize green or orange. If the display turns red, the power supply
unit must be plugged in immediately. Only in this case the aspiration can
be continued without any interruption.
Operation
EN
26