4.1.6 Cleaning / disinfection of the holder for external canister "Bag"

Clean the surfaces of the holder for external canister "Bag" regularly and
disinfect them daily.
The holder for external canister "Bag" can be wiped with a damp,
lint-free cloth.
Follow the instructions in section 4.1.1 for wipe disinfection.
Repeated cleaning and disinfection procedures may result in minor discolorations of
the plastic component of the holder for external canister "Bag". However these do
not affect the function of the device.
Löwenstein Medical GmbH & Co. KG recommends to replace the holder for external
canister "Bag" every 12 month, if it is used and disinfected frequently and for
each new patient.

4.1.7 Cleaning / disposal of the suction tube

1. Dispose of the suction tube properly before the device is used by a new patient!
EN
2. If the device is used for one patient only, rinse the suction tube with clear water
(max. 0.5 l) using the rinsing bottle after each aspiration and place it daily in
33
the disinfectant solution recommended by Löwenstein Medical GmbH & Co. KG
for immersion disinfection (see section 4.1.1).
3. Afterwards, rinse the suction tube thoroughly with clear water and let it dry.
Even when the suction tube is used for one patient only, Löwenstein Medical GmbH
& Co. KG, recommends to replace the suction tube at least every 4 weeks and
for each new patient since the material may become discolored and brittle due to
frequent applications!

4.1.8 Cleaning and disinfection of the rinsing bottle (250 ml)

1. Remove the lid of the rinsing bottle and rinse both parts under running water.
2. Immerse the rinsing bottle and the lid in disinfectant solution, considering the
specific concentration quantity.
3. Afterwards, rinse the components thoroughly and let them dry.
4. Place the lid back on the rinsing bottle.
Löwenstein Medical GmbH & Co. KG recommends to replace the rinsing bottle
every 4 weeks, if it is used and disinfected frequently and for each new patient!