Install the removable terminal blocks onto the correct connectors on the terminal.
2.6.2.
Junction Box
Make certain the DIN enclosure and load cell junction board are securely attached inside the
junction box enclosure.
Make sure the gasket is in good condition and properly positioned.
Place the cover over the junction box enclosure.
2.6.2.1.
Original Version
This version secures the front cover to the rear enclosure with eight bolts from the rear. Four of the
eight screws are shown in Figure 2-75.
Insert the eight bolts through the holes in the cover and start each bolt into the threaded receivers in
the rear enclosure.
After all bolts have been started, tighten each bolt to 1.5 N-m (12 lbf-in). Version mise à jour.
2-46
METTLER TOLEDO IND131/IND131x/IND331/IND331xx Installation Manual