2.4.5.5.

Allen Bradley RIO

Connections to the Remote IO option board (Figure 2-55) are made using a three-pin terminal
connector on the RIO option.
The connection should be wired as shown in Figure 2-56.
The part number for the Remote IO cable is Belden 9463 and is sometimes referred to as "Blue
Hose" cable. The METTLER TOLEDO part number for this cable is 64056504.
2.4.5.6.

CC-Link Interface

The optional CC-Link board (Figure 2-57) connects to the network via shielded twisted pair cables.
The module's address is set in software.
64067480
|
11
|
05/2018
Figure 2-55: A-B RIO Option Board
NOTES:
1. CONNECTION WITH TWIN AXIAL CABLE (BLUE HOSE) SHOWN.
2. REFER TO ALLEN-BRADLEY REMOTE I/O DOCUMENTATION FOR TERMINATION
RESISTOR AND OTHER CONSIDERATIONS.
3. WIRE SIZE: 14 AWG (2.0 mm2) MAXIMUM
22 AWG (0.3 mm2) MINIMUM.
Figure 2-56: RIO Connection
THE CC-LINK PLC OPTION (#30059622) MUST NOT BE USED IN A TERMINAL INSTALLED IN
AN AREA CLASSIFIED AS DIVISION 2 OR ZONE 2/22. FAILURE TO COMPLY WITH THIS
WARNING COULD RESULT IN BODILY HARM AND/OR PROPERTY DAMAGE.
METTLER TOLEDO IND131/IND131x/IND331/IND331xx Installation Manual
WARNING
2-35