The summing board can then be placed on the bottom of the enclosure for wiring.
Prepare the ends of each of the load cell cables as shown in Figure 2-40 and Table 2-6. Make sure
the shield is connected to the gland as depicted in Figure 2-28.
Insert the load cell cables into the correct gland as identified in Figure 2-40
2-24
METTLER TOLEDO IND131/IND131x/IND331/IND331xx Installation Manual
Figure 2-37: Removing the Summing Board
A
Figure 2-38: Load Cell Cable Preparation
Table 2-5: Load Cell Cable Length
Cable
Shield Length (A)
Load Cell 1
25mm (1 in.)
Load Cell 2
25mm (1 in.)
Load Cell 3
25mm (1 in.)
Load Cell 4
25mm (1 in.)
B
Stripped Cable Length (B)
127mm (5.0 in.)
127mm (5.0 in.)
190mm (7.5 in.)
190mm (7.5 in.)
Wire tie for load
cell cable harness