•
•
•
2.4.
Main Board Wiring Connections
The wiring connector positions for the DIN, Junction Box and Panel-mount versions are shown in
Figure 2-30. The harsh enclosure is shown in Figure 2-31.
2-18
METTLER TOLEDO IND131/IND131x/IND331/IND331xx Installation Manual
Depending upon the diameter of the cable to be installed in a 16mm gland, select one of the
two different sized rubber grommets (if required) to properly seal around the cable. There are
no grommets for the 12mm glands.
Grommet
None
Larger size hole
Smaller size hole
When making cable terminations inside the enclosure, ensure that the cable length from the
terminal strip/connector to the terminal housing is sufficient so that no strain is placed on the
connector assembly when the housing is in the fully open position.
After making the wiring connections as described in the next section, ensure the nut on the
cable gland is tightened properly to seal around the cable. Ensure that this seal is watertight.
7-position
Analog Load
Cell interface
COM1 serial
port
AC power
connector
Figure 2-30: DIN, J-Box and Panel-mount Connections
Table 2-2: 16mm Grommet Cable Sizes
Cable Diameter
7–10 mm (0.28–0.39 in.)
5– 6 mm (0.20–0.24 in.)
3–4 mm (0.12–0.16 in.)
64067480
|
11
|
05/2018