HBM PMX Manuel D'emploi page 124

Masquer les pouces Voir aussi pour PMX:
Table des Matières

Publicité

124
MRG
This command assigns a measurement rate group to a selected channel or
subchannel (PCS / SPS). Up to 3 synchronous measurement rate groups are
supported. The measured values of the various groups are stored in separate
FIFO buffers and must be read out separately over the interface.
Syntax:
MRG p1,p2,p3 (x)
Parameter:
p1: 0..2 measurement rate group
p2: 0..2 measurement rate group (optional)
p3: 0..2 measurement rate group (optional)
Note:
Response depends on SRB command.
MRG?
Syntax:
MRG? (x)
Response:
q1(y)
q1: measurement rate group
Example: MrgOfSubSignal11 : MrgOfSubSignal12 : MrgOfSubSignal21 :
MrgOfSubSignal22 ...
ICR
This command is used to set up (only) one measurement rate per group. A
second measurement rate in a group is not supported.
Syntax:
ICR p1, p2(x)
Parameter:
p1:
p2:
If parameter p2 is omitted, the command affects measurement rate group 0.
Measurement Rate Group
Output the measurement signal selection for the channels to be re-
corded
Measurement Rate Group Query
Internal Channel Recordingrate
measurement rate 1, see rate list below
0, 1, 2 ; measurement rate group
Commmands
A3257−2.0 en/de/fr