Access Codes; Block Levels - Burkert SE 56 Manuel D'instructions

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

ACCESS CODES

Some functions in the transmitter are
enabled by the access codes. The
information within this manual is related to all
the functions available with L2 level. All the
functions available through higher level are
protected and reserved to the service.
Description of the L2 access code
(Menu "11 Internal data" pos. 11.1)
with code L2 = 00000 you disable the
request of code
NOTE: the availability of the functions is related to
the selected block
* with L2 customised (freely chosen by
the user) you can proceed programming
all the functions up to L2 security level,
entering the code itself whenever you
need to enter the Main menu
*ATTENTION:
take note very carefully of the
customised code you have chosen, since there is
no way for the user to retrieve it if it is forgotten

BLOCK LEVELS

The block level enables or disables the access to the
functions of the transmitter.
The available levels of block are the following:
(Menu "11 Internal data" pos. 11.2)
Level 0: completely disables access to the functions.
You can perform the following functions through the
keyboard:
Changing the display mode
Dosing Start/stop
Data printing
Level 1: Enables access to the following functions:
Totalise re-setting
Dosing functions modifications
Level 2: Enables access to the following functions:
Quick start menu
Scale (full enabling)
Display (partial enabling)
Diagnostics (partial enabling)
Level 3: Enables access to all the functions of level 2
as well as maintenance features
FACTORY PRE-SETTINGS
ACCESS CODES
The transmitter is delivered with access
code L2:
and with the "Quick start menu" enable.
Press the "ENTER" key to access to the
"Quick start menu" from any of the
visualization pages
The "Quick start menu" is enabled by
function 8.6 (display menu); from the "Quick
start menu" the functions may be set
without entering any access code (see
example 1 on page 24).
The last function allows to access the main
menu.
FACTORY PRE-SETTINGS
BLOCK LEVELS
The transmitter is delivered with the following block
level:
If for several reasons you need to change the block
level, follow the steps:
Set the dip switch to OFF position
Access to the function "
(main menu)
Choose the desired level of block
To enable the selected block level place the DIP switch
to ON position
When the Dip-switches are set to "OFF", all functions
are available.
(OFF)
26
11111
3
Block level
" of menu 11
LOCK
(ON)
Dip
switch
SE 56

Publicité

Table des Matières
loading