EN
14.4
Run-up and run-down parameters
The run-up and run-down parameters that have been set are displayed.
x: 1-9 = run-up step, t = run-up time
y
x
y: R1-R9, B1-B9 = braking step, R0 = unbraked run-down, t = run-down time, n
energisation speed
PROFIL
14.4.1
Run-up step
Press the
key repeatedly until the parameter
inversely.
Set the desired step by means of the control knob
14.4.2
Run-up time
Press the
key repeatedly until the parameter
Set the desired value by means of the control knob
If a run-up time is set that is longer than the running time, the centrifugation run will end before the set speed is
reached.
14.4.3
Braking step
Press the
key repeatedly until the parameter
inversely.
Set the desired step by means of the control knob
B-braking steps can only be set for special rotors.
14.4.4
Run-down time
If a brake de-energisation speed has been set, it is not possible to set a run-down time.
Press the
key repeatedly until the parameter
Set the desired value by means of the control knob
14.4.5
Brake de-energisation speed
Press the
key repeatedly until the parameter n
Set the desired value by means of the control knob
14.5
Radius/Temperature
14.5.1
Radius
Press the
key repeatedly until the parameter r/mm is displayed and the input field is shown inversely.
Set the desired value by means of the control knob
The value of the RCF/RZB adapts automatically when the radius is changed; this is shown by the indication flashing.
14.5.2
Temperature
Press the
key repeatedly until the parameter T/°C is displayed and the input field is shown inversely.
Set the desired value by means of the control knob
14.6
Automatic buffer
The buffer comprises the program locations "----" and 90 to 99. Every time a centrifugation run is started, modified
centrifugation data are automatically stored in program location "----". The modified centrifugation data for the last 11
centrifugation runs are stored in the buffer and can be called up (see Chapter "Calling up programs").
48/137
or
is displayed. The input field will be displayed
t
1-9
.
is displayed. The input field will be displayed inversely.
min:sec
t
.
or
is displayed. The input field will be displayed
t
0-9
.
is displayed. The input field will be displayed inversely.
t
min:sec
.
(•)
/RPM is displayed. The input field will be displayed inversely.
.
.
.
(•)
= brake de-