Protection fuse
Pwr supply (24 Vdc):
Protection fuse
Compliance with:
Power cable
Serial communication (Navigator
kit)
Pump Input
Lubricant
Installation category
Pollution degree
Storage temperature
Weight kg (lbs):
Kit:
Controller:
* (According to standard DIN 28 428, the base pressure is that
measured in a leak-free test dome, 48 hours after the completion of
test dome bake-out, with a Turbopump fitted with a ConFlat flange
and using the recommended pre-vacuum pump).
When the TV 301 Navigator has been stored at a temperature less than 5°C,
NOTE
wait until the system has reached the above mentioned temperature.
TV 301 Navigator User Manual / 87-900-946-01 (N)
Input voltage:
Max input pwr:
Stand-by pwr:
Max oper. power:
Technical Information
Technical Specification
4 A
24 Vdc
200 W
10 W
150 W with water or air cooling
8 A
UNI EN 292-1
UNI EN 292-2
EN-CENELEC 55011
IEC 1000-4-2 (ex 801-2)
IEC 1000-4-3 (ex 801-3)
IEC 1000-4-4 (ex 801-4)
EN 61010-1 (IEC 1010-1)
EN 1012-2
With European or NEMA plug 3 meters long
(optional)
RS232 cable with a 9-pin D type male connector
and a 9-pin D type female connector, and Navigator
software (optional)
75 Vac, three phase, 963 Hz
permanent lubrication
II
2
- 20 ° C to + 70 ° C
ISO 100 and 160: 4.5 (10), CFF 6" and 8": 8 (17.6)
1 (2.2)
17
227/284