Signal Description

Start/Stop: input signal to start or stop the pump. With the supplied
cover connector the START/STOP (+) signal is connected to the +24
Vdc pin and the START/STOP (-) signal to the GROUND pin: in this
condition the pump automatically starts as soon as the controller
recognises the input supply ("Plug & Pump").
Interlock: input signal to control the pump rotation. With the
supplied cover connector the interlock (+) signal is connected to the
+24 Vdc pin and the interlock (-) signal to the GROUND pin.
Soft start: this input is used to provide a "soft start" to the pump; in
this condition the ramp-up time could be up to 45 min.
TV 301 Navigator User Manual / 87-900-946-01 (N)
Signal characteristics and use
SIGNAL NAME
Start/Stop (+)
Start/Stop (-)
Interlock (+)
Interlock (-)
Speed setting (+)
Speed setting (-)
Soft start (+)
Soft start (-)
+ 24 vdc
Spare
Programmable set point
Spare
Fault output
Programmable analog signal (+)
• Ground
• Programmable analog signal (-)
Technical Information
Interconnections
IN-/OUTPUT
In
In
In
In
In
In
In
In
Out
Out
Out
Out
Out