Switching on and Use of Turbo-V 1K-G

The following instructions apply to the Turbo-V 1K-G used togheter
with Navigator controller (model 9698978M005). Please refer to the
relative control unit manual for use with other controller model
(9699454M002).
To switch on the Turbo-V 1K-G it is necessary to supply the mains.
The integrated controller automatically recognizes the interlock and
start signals presence and start up the pump.
The first pump start up is in "Soft Start" mode. When the start up
cycle is finished, the "Soft Start" mode automatically is disabled, and
the following start ups are without the "Soft Start" mode. To re-
enable the "Soft Start" mode it must be activated by the suitable
software command (see the paragraph "RS 232/RS 485
COMMUNICATION DESCRIPTION" in the appendix "Technical
Information").
The green LED located on the Navigator controller base front panel
indicates with its flashing frequency the system operating conditions:
Turbo-V 1K-G User Manual/ 87-901-025-01 (A)
with no flashing: the pump is normally rotating;
slowly flashing (period of about 400 ms): the system is in ramp,
or in braking, or in Stop, or in "Waiting for interlock" status;
fast flashing (period of about 200 ms): error condition.