Cooling may be carried out either through an open circuit with
eventual discharge of the water, or using a closed circuit cooling
system.
The water temperature must be between +10°C and +20°C, with an
inlet pressure between 3 and 5 bar. This allows a flow of about 200
l/h.
The water electrical conductance must be ≤500 µs/cm. When the conductance
NOTE
is higher, in closed water circuit, the use of up to 20% of Ethyl-Glycole is
suggested.
Turbo-V 1K-G User Manual/ 87-901-025-01 (A)
Technical Information
Water Cooling Kit Installation
17