Instructions for Use
Pump Fixing
Pump Fixing
WARNING!
If a rotor failure occurs, the connection of the pump to the system could be
subjected to a significant torque. If the connection is not sufficient to withstand
that torque, the pump could detach from the system or the motor housing could
detach from the pump envelope. In this case metal fragments could be
projected from the pump or system, which could cause serious injury or death
and/or damage to surrounding equipment.
Fix the Turbo-V 1K-G in a stable position mounting the inlet flange of
the turbopump to the system counter-flange, with a connection
capable of withstanding a torque of 8900 Nm around its axis.
The following table shows, for each ISO-F flange, the necessary
number of screws and the relevant fixing torque.
Table 1
Flange
ISO 160 F
ISO 200 F
The class of the steel screws for "F" flange must be > 8.8
The Turbo-V 1K-G can be installed in any position.
The Turbo-V 1K-G cannot be fixed by means of its base.
NOTE
238/304
N.
8
12
Turbo-V 1K-G User Manual / 87-901-025-01 (A)
Fixing Torque
25 Nm
25 Nm