17
Technical Information
Water Cooling Kit Installation
The assembled kit must be screwed into the suitable holes of the
pump body with a recommended closing torque of 5 Nm.
Two 6 mm (1/4") internal diameter rubber or plastic hoses from the
water supply must be fitted to the two nozzles.
These hoses must be held on the respective nozzles using hose clips to avoid
NOTE
that the tube(s) gets loose or disconnected during operation.
282/304
Turbo-V 1K-G User Manual / 87-901-025-01 (A)