Switching On And Use Of Tv 902 Navigator - Agilent Technologies TV 902 Navigator Notice De Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 41

Switching on and Use of TV 902 Navigator

To switch on the TV 902 Navigator it is necessary to supply the
mains. The integrated controller automatically recognizes the
interlock and start signals presence and start up the pump.
"Soft Start" mode is provided to start the pump after a protracted
stop. This allows a better grease distribution in the bearings. To
enable the "Soft Start" mode it must be activated by the suitable
software command (see the paragraph "RS 232/485
COMMUNICATION DESCRIPTION" in the appendix "Technical
Information").
The green LED located on the TV 902 Navigator base front panel
indicates with its flashing frequency the system operating conditions:
TV 902 Navigator User Manual / 87-900-999-01 (B)
with no flashing: the pump is normally rotating;
slowly flashing (period of about 400 ms): the system is in ramp,
or in braking, or in Stop, or in "Waiting for interlock" status;
fast flashing (period of about 200 ms): error condition.
Instructions for Use
Use
233/276
16

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

8698933r001Sq 337

Table des Matières