OPERATING THE OPTIONAL TWO-STAGE PERMANENTLY ATTACHED I.V. POLE

Note: The two-stage permanently attached I.V. pole is an option and may have been installed at either the head, foot
or both ends of the stretcher. The choice was made at the time that the stretcher was purchased.
To use the two-stage permanently attached I.V. pole:
1.
Lift and pivot the pole from the storage position and push down until it is locked into the receptacle.
2.
To raise the height of the pole, pull up on the telescoping portion (A) until it locks into place at its fully raised
position.
3.
Rotate the I.V. hangers (B) to desired position and hang the I.V. bags.
4.
To lower the I.V. pole, turn the latch (C) until section (A) lowers.
B
A
C
CAUTION
To avoid damage, the weight of the I.V. bags should not exceed 40 pounds.
To avoid damage while transporting the stretcher, verify that the I.V. pole is at a low enough height to allow it to safely
pass through door openings and under light fixtures.
www.stryker.com
Operation Guide
Figure 14: I.V. Pole
1079-009-005 REV E
Figure 15: I.V. Pole Latch Detail
Return To Table of Contents