5 MAINTENANCE

5.1 Maintenance
In correct installation conditions the flow sensors and totalizers are maintenance-free. If
contamination or clogging should occur during operation, the transmitter (paddle-wheel,
bearing) can be cleaned with water or another appropriate cleaning agent.
The message "ERROR" on the display indicates that calibration data has been lost. By
pressing ENTER, the user access to operation menu but the device works with the factory
settings. The transmitter must be re-calibrated. If this message appears more often, please
return the product to the factory.
When the batteries become low, the display blinks (flow display and totalizer) but the
device still works. A replacement of batteries muss be planed. The message "LOW BATT"
appears only when the batteries are too low to ensure function of the device.
5.2 Configuration of flow indicator-totalizer SE35 at delivery
Language:
Unit flow:
Unit totalizer:
Decimal point:2
K-factor:
Filter:

5.3 Spare Parts List

Position
Specification
1
Sensor housing with coil measuring element
2+3
Set with 2 M20x1.5 screw plugs + 2 flat seals
4
Cover, scew, folio and electronic board
Set incl. 8 FLOW folios
Instruction manual Fitting Typ S030
English
L/s
L
46,6
Filter 2
1
3
E-14-
FLOW TOTALIZER 8035
2
4
Order code
425247
444705
425433
553191
426107