4 OPERATION

4.3.3 K-Factor

The K-factor of the fitting is entered in this menu (refer to the instruction manual of fitting
S030). The "Teach in" function allows the practical determination of the specific K-factor.
The user only needs to run a known volume through his system.
Example: In order to determine a volume as accurately as possible, the user will fill a tank
of 100 liters. When the message "TEACH YES" appears, he presses the ENTER key to start
the measuring procedure. The message "FILL END" (end of filling) will appear. He then
switches on a pump or opens a valve. As soon as the tank is empty, he switches off the
pump or closes the valve. Pressing ENTER ends the measurement. The user will then be
asked to enter the volume (100 liters). The calculated K-factor is displayed after
confimation.
Note: The device uses the last K-factor entered or determined .
FACT. K
ENTER
FILTER
FILTER
TEACH N
ENTER
ENTER
TEACH O
ENTER
Start of measurement.
ENTER
E-12-
FLOW TOTALIZER 8035
K = 000.00
Entering of K-factor as
specified in the charts.
0......9
K = 117.60
END COMP
ENTER
End of
measurement.
Entering the measured volume.
Same unit as for flow.
K = 79.30
ENTER
Indication of calculated
K-factor.
0000.0 L
0......9
0100.0 L