Battery Calibration; Calibrating Nimh Rechargeable Battery; Storage Of The Rechargeable Battery - Medela Clario Mode D'emploi

Masquer les pouces Voir aussi pour Clario:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 66

12 Battery calibration

(AC/DC version only)
CAUTIONS
Calibrate the rechargeable battery before first use and after storage duration > 2 months�
Replace battery if the calibration is not successful�
Calibrating NiMH
rechargeable battery
Storage of the
rechargeable battery
1�
Make sure that the rechargeable battery is fully charged�
2�
Pull out the mains plug from the fixed mains socket�
3�
4�
Let the pump run until the rechargeable battery is empty
and the pump comes to a stop�
5�
Connect the Clario to a fixed mains socket�
The rechargeable battery is being charged�
6�
Wait until the charging indicator goes out�
7�
Repeat step 1 to 6� The Clario AC/DC is then ready to use�
In order to keep the self-discharge at a minimum, store the
Clario and the replacement rechargeable batteries at tempera-
tures below 25 °C (77 °F)� Repeat the calibration process every
60–90 days�
3�1 Switch on the Clario� The
pump will start running�
Battery calibration
| 23

Publicité

Table des Matières
loading

Table des Matières