Electromagnetic Immunity - Medela Clario Mode D'emploi

Masquer les pouces Voir aussi pour Clario:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 66
Electromagnetic Compatibility (EMC, IEC 60601-1-2:2007, Table 2)

Electromagnetic Immunity

Clario is intended for use in the electromagnetic environment specified below� The cus-
tomer or the user of Clario should assure that they are used in such environment�
Immunity Tests
IEC 60601
test level
Electrostatic
± 6 kV contact
Discharge (ESD)
± 8 kV air
IEC 61000-4-2
Electrical fast
± 2 kV
transient/burst
for power
IEC 61000-4-4
supply lines
± 1 kV
for input /
output lines
Surge
± 1 kV
IEC 61000-4-5
differential mode
± 2 kV
common mode
<5 % U
Voltage dips,
(>95 % dip in U
short interrup-
tions and voltage
for 0�5 cycle
variations on
40 % U
power supply
(60 % dip in U
input lines
IEC 61000-4-11
for 5 cycles
70 % U
(30 % dip in U
for 25 cycles
<5 % U
(>95 % dip in U
for 5 sec
Power frequency
3 A/m
(50 / 60 Hz)
magnetic field
IEC 61000-4-8
Compliance
level
± 6 kV contact
± 8 kV air
± 2 kV
for power
supply lines
± 1 kV
for input /
output lines
± 1 kV
differential mode
± 2 kV
common mode
<5 % U
T
T
(>95 % dip in U
)
T
for 0�5 cycle
40 % U
T
T
(60 % dip in U
)
T
for 5 cycles
70 % U
T
T
(30 % dip in U
)
T
for 25 cycles
<5 % U
T
T
(>95 % dip in U
)
T
for 5 sec
3 A/m
Electromagnetic
environment – guidance
Floors should be wood, concrete or
ceramic tile� If floors are covered with
synthetic material, the relative humidity
should be at least 30 %�
Mains power quality should be that
of a typical commercial or hospital
environment�
Mains power quality should be that
of a typical commercial or hospital
environment�
Mains power quality should be that of
)
a typical commercial or hospital envi-
T
ronment� If the user of Clario requires
continued operation during power
mains interruptions, it is recommended
)
that Clario be powered from an unin-
T
terruptible power supply or a battery�
)
T
)
T
Power frequency magnetic fields
should be at levels of a typical com-
mercial or hospital environment�
Technical documentation
| 323

Publicité

Table des Matières
loading

Table des Matières