6 Preparation for use
WARNINGS
For use only by medically trained persons who have been adequately trained in suction
procedures and in the use of aspirators� Wear gloves for all operations�
CAUTIONS
– The Clario must remain in an upright position during use�
– Sterile accessories must be checked on the integrity of the packaging before use�
– Non sterile and reusable accessories must be cleaned, disinfected and / or sterilized
according to the Medela cleaning guide (product code 200�2391)�
6.1
Checks before use
6.2
Assembly of the
canister
– Check the Clario system before use for damage of the
power cord or plug, obvious device damage or safety
defects and proper functioning of the device.
– Check for completeness and general condition of the
Clario delivery package.
– Only for AC / DC versions: make sure that the recharge-
able battery is calibrated according to chapter 12.
– Check all accessories prior to use:
– Canister and lid for cracks, brittle and flawed spots.
Replace if necessary.
– Tubing for cracks, brittle areas and that connectors are
firmly attached. Replace if necessary.
1�
2�
1�1 Attach the floater to the
canister�
2�1 Attach the lid to the
canister�
Mind the markings on the
front of the canister and
the lid�
Preparation for use
| 13