11 Battery data and test

(AC / DC version only)
CAUTIONS
Only charge the rechargeable battery when the charging indicator blinks slowly�
The rechargeable battery is charged during mains operation�

Duration of pumping

If the battery is fully charged, the operating time at maximum vacuum is approximately
> 50 minutes�

Charging time (charged up to 100 %)

Approx� 5 hours�

Battery status

Testing the
rechargeable battery
22 |
Battery data and test
not lit
Battery is fully charged
lit up
– Battery is being charged
– Pump is plugged in
blinks slowly
Battery is soon empty
blinks rapidly
Battery is defective
1�
Make sure that the rechargeable battery is fully charged�
2�
Pull out the mains plug from the fixed mains socket�
3�
4�
5� Measure the time until the
pump comes to a stop�
The battery is ok, when
the the running time is
> 50 minutes�
3�1 Set vacuum regulator to
position «max�»�
4�1 Switch on the Clario� The
pump will start running�
If the running time is
< 50 minutes, repeat test
or calibrate the battery
(see next page).