AC/DC versions
Vll
Indicators
during battery operation
During mains operation
E P !
B E
E P !
B E
E P !
B E
Performance of rechargeable battery
Duration of pumping
Charging time (charged up to 100%)
Rechargeable battery is charged during mains operation.
Testing of rechargeable battery
1. Pull out mains plug
2. Set required vacuum
3. Plug the suction port
4. Switch on the Vario
5. Measure the running time
6. Compare the running time
ok?
not OK
• Repeat the test
• Calibrating the rechargeable battery
NiMH rechargeable battery
Before first use
Operation with
rechargeable
battery
Storage
6
EN
Visual
does not light up
blinks slowly
lights up
does not light up
E P !
blinks rapidly
B E
E P !
B E
E P !
B E
OK, if the above values are reached
Calibrating the rechargeable battery
A Completely discharge the rechargeable battery
• Disconnect the mains cable
• Switch on the Vario and allow to run until it
comes to a standstill
B Charge the Vario
• Plug in the mains cable – the rechargeable
battery is being charged
• Wait until the charging indicator goes out
*
Only charge the rechargeable battery when the acoustic signal
sounds and the charging indicator blinks slowly
In order to keep the self-discharge at a minimum, store the Vario
and the replacement rechargeable batteries at temperatures below
25°C (77°F). Repeat Steps A and B every 60-90 days
Acoustic
rechargeable battery is fully charged
rechargeable battery soon empty
rechargeable battery is being charged
rechargeable battery is fully charged
rechargeable battery is defective
low
medium
vacuum
vacuum
~120 Mins.
~ 30 Mins.
~ 5 Hrs.
~ 5 Hrs.
low
medium
vacuum
vacuum
-90 mbar
- 9 kPa
-55 kPa
-68 mmHg
-413 mmHg
> 120 Mins.
> 30 Mins.
high
vacuum
~ 30 Mins.
~ 5 Hrs.
high
vacuum
-55 kPa
-413 mmHg
> 30 Mins.
Repeat this
process
once
The Vario is