Troubleshooting; Motor Not Running; Pump Is Too Loud; Insufficient Vacuum - Medela Clario Mode D'emploi

Masquer les pouces Voir aussi pour Clario:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 66

13 Troubleshooting

Motor not running

Check if:
– the Clario is switched on�
– the mains voltage is correct and if the mains plug is inserted correctly into the fixed mains
socket�
– the internal battery is charged (only AC / DC version under battery operation)�

Pump is too loud

Check the position and seal of the valve discs inside the SafetyChamber�

Insufficient vacuum

Check if:
– the vacuum regulator is set correctly�
– the tubings are not defectiv e or broken� If necessary, replace�
– all plug-in connections are tight�
– the suction canister / lid have no cracks, brittle areas, discolouration� If necessary, replace�
– the seal of the valve discs inside the SafetyChamber is tight�
– the membrane is clean and has no tears�

Running time from battery insufficient

Calibrate the rechargeable battery� See chapter 12�

No LED lit (AC / DC version only)

The rechargeable battery is fully charged�

LED lit up (AC / DC version only)

– The rechargeable battery is being charged�
– Pump is plugged in�

LED blinks slowly (AC / DC version only)

The rechargeable battery is soon empty� The battery power is sufficient for about 10 minutes
of pumping�

LED blinks rapidly (AC / DC version only)

The rechargeable battery is defective� For replacing the defective battery see chapter 10�
24 |
Troubleshooting

Publicité

Table des Matières
loading

Table des Matières