8. Technical Data
General
Safe working load
Maximum patient weight
Product weight (approx.)
Audible noise
Operating conditions
Temperature
Relative humidity
Atmospheric pressure
Electrical data
Power input
Duty cycle
Safety standards USA/Canada
Other Safety standards
Electric shock protection
EMC
Potential equalisation terminal
Liquid ingress protection
Backup battery
58
250kg (550 lbs)
185kg (407 lbs)
180kg (396 lbs)
50dB approx.
5°C to 40°C (41° F to 104 °F)
20% to 90% at 30°C (86 °F), non-
condensing
700hPa to 1060hPa
3A max. at 230V a.c. 50/60Hz
3A max. at 230V a.c. 60Hz (KSA)
5.8A max. at 120V a.c. 60Hz
10% (2 min. on, 18 min. off)
Medical equipment classified in
accordance with UL 60601-1
CAN/CSA C22.2 No.601.1M90
IEC 60601-2-38 and
ANSI/AAMI ES60601-1:2005
CAN/CSA C22.2 No.60601-1:08
IEC 60601-2-52:2009
IEC 60601-1:1988+A1+A2
IEC 60601-2-38:1996+A1:1999
IEC 60601-1:2005 &
IEC 60601-2-52:2009
Class I
Type B
Complies with EN 60601-1-2:2002 and
2007
Complies with EN 60601-1:1990 and 2006
IPX4
2 x 12V series connected, sealed,
rechargeable lead/acid gel, 1.2Ah