Arjohuntleigh Enterprise 9600X Notice D'utilisation page 60

Masquer les pouces Voir aussi pour Enterprise 9600X:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

WARNING
The procedures below must be carried out by suitably trained and
qualified personnel. Failure to do so may result in injury or an
unsafe product.
Actions to be done by qualified personnel
Check that the bedstripper (linen shelf) remains in its closed
position when maximum foot down tilt is applied
Check that the bed operates correctly using the backup battery
as described in the section "Battery test" on page 53.
Check operation the of castors, paying special attention to
braking and steering functions
Check that the bed extension locks securely in all three positions
Examine the power supply cord and mains plug; if damaged,
replace the complete assembly; do not use a rewireable plug
Examine all accessible flexible cables for damage and
deterioration
Check all accessible nuts, bolts and other fasteners are present
and correctly tightened
Check any accessories fitted to the bed, paying particular
attention to fasteners and moving parts
Weighing system
WARNING
The weighing system must be re-calibrated by a qualified service
engineer before the date of expiry shown on the bed
(Fig 27, Item 1). If no weighing calibration label is present, the
weighing system must be re-calibrated.
Make sure the bed is re-calibrated every 12 months.
52
Fig. 27 - Weighing Calibration Expiry Label
1
Yearly