Symptom
TV/lighting controls not
working
Red brake light on
SafeSet indicator panel
but the brakes are in the
on position
Red side rail light on
SafeSet indicator panel
but the side rails are in the
raised position
No lights on SafeSet
indicator panel
SafeSet indicator panel
lights always bright
Product lifetime
The lifetime of this equipment is typically ten (10) years.
"Lifetime" is defined as the period during which the product will
maintain the specified performance and safety, provided it has
been maintained and operated in conditions of normal use in
accordance with the requirements in these instructions.
56
Possible Cause
Nurse call cable not
connected or wrong type
Brake sensor failure
Side rail sensor failure
Bed disconnected from
electricity supply
Light sensor obscured or
failed.
Action
Check cable is of correct
type and properly
connected
Replace brake sensor
Replace side rail sensor
Bed running in power
conservation mode.
Reconnect electricity
supply and swipe fob to
reactivate SafeSet
indicator panel.
Remove any items
obscuring the SafeSet
indicator panel light
sensor. If no improvement
call an ArjoHuntleigh
approved service
engineer