Error codes are shown on the display. They are used to
Error Codes
indicate a problem with the weighing system; this may be due
to an operator error or a possible fault condition.
The table below shows common error codes with their
meanings and possible causes.
Display
E001
E002
E003
E102
If the display shows an error code other than those given above,
refer to the product service manual or contact an approved
ArjoHuntleigh service agent.
When operating the controls for backrest or tilt, the weighing
Angle
indication
system display shows the approximate angle in degrees of the
selected function.
Angles are shown relative to the floor, so the displayed values
of backrest angle and thigh angle change when the mattress
platform is tilted.
The tilt angle is shown as a positive value for head down tilt
and a negative value for foot down tilt.
40
Cause
Zero on battery power
Auto Compensation
weight decrease more
than 100kg (220 lbs.)
Auto Compensation
weight increase more
than 100kg (220 lbs.)
A stable weight reading
was not obtained within
10 seconds
Solution
Connect bed to mains supply and
re-initialise the weighing system -
see page 38
Replace weight removed from
bed
Remove added weight from bed
Refer to the points listed in the
section "Accuracy" on page 37