Result Number
none
ANALYZER LIFE
REMAINING
XXX READS
none
REMAINING
Flashing Plug icon
none
WALK AWAY MODE
REMOVE TEST DEVICE
none
WHEN PROMPTED
TEST DEVICE
none
WAS REMOVED
REPEAT ASSAY
Message
Possible Cause(s)
The shelf or operational life
of the BD Veritor Plus
Analyzer is nearing
expiration. The maximum
XXX DAYS
shelf-life of the BD Veritor
Plus Analyzer from the date
of manufacture is 34 months
(regardless of time of first
use). The maximum number
of months from first use is
24 months.
The BD Veritor Plus Analyzer
is nearing its usage
expiration. The maximum
number of tests performed by
a BD Veritor Plus Analyzer is
3,500 runs.
Failure of the internal battery.
Test device inserted at the
incorrect workflow step
INSERT
during Walk Away operation
mode.
Test device was removed
before the test run was
complete during Walk Away
operation mode.
Corrective Action(s)
Contact your distributor to order a new BD Veritor Plus
Analyzer.
The duration of the message is two seconds. To clear the
message immediately, insert a test device to begin a new
run.
Contact your distributor to order a new BD Veritor Plus
Analyzer.
The duration of the message is two seconds. To clear the
message immediately, insert a test device to begin a new
run.
Contact BD.
Note: the Analyzer may continue to be used while
plugged into a facility power source.
Current run aborted. Remove test device and
immediately restart the Walk Away test sequence. Insert
the test device after the display window prompt to do so.
Current run aborted. The test must be repeated with a
new test device because the elapsed development time
of the assay is unknown.